WWW.MEASURETEK.NET PS-102 User/Technical Manual Contents subject to change without notice Version 1.0 08/2013
Table1: use Kg as calibration unit: Display division value in different weight unit that can be used Calibration division value kg lb
0.02 lb 0.01 kg 0.02 lb 0.5 oz 0.2 lb 0.1 kg 0.2 lb 5 oz 2 lb 1 kg 2 lb Not available 20 lb 10 kg 20 lb Not available 0.0005 lb 0.0002 k
7. User Parameters Setup 1. In normal weighing mode, press UNIT and ZERO/ON/OFF until “USer” is shown to enter into the user parameters
4=continuously output gross, tare, net weight, unit and status data, data has not been received; 5=output weight, unit and status data one time when
The following details contain more information when U5 is set to 7: a) The baud rate and data format is set by U6 and U7. Responses to serial comman
Response: <LF>H1H2H3<CR><ETX> Zero function is activated and then it returns to current scale status, similar to pressing ZERO key.
4 always 1 always 1 always 1 5 always 1 always 1 always 1 6 always 0 always 1 always 0 7 parity Parity parity 8. Symbol Definitions
0﹉﹉﹉ Indication is out of key zero range Reduce the weight on platform until the indication is within the key zero range. 0﹍﹍﹍ Weight reading below P
PS‐102 www.measuretek.net .16 11.One Year Limited Warranty MeasureTek products covered in this manual are guaranteed to be free from defects in
TABLE OF CONTENTS 1. INTRODUCTION and INSTALLATION ...
1. INTRODUCTION and INSTALLATION General and Safety Information Risk of Electrical Shock: Disconnect all power sources before making cable connec
Contents Indicator Base AC120V/DC6V 500mA UL adapter 6’ USB Cable Quick Guide & Technical Manual Connection types 4 pins socket use
Indicator Display >0<- Scale is zeroed and gross weight is 0, tare is 0. Net - Display reading is net weight; tare is not 0. lb, kg, oz
4. To change the weight unit of measure, press the UNIT key. 5. To send data to another device via the serial port, press the PRINT/HOLD key. 6. T
displayed. The default standard weight is 100% FS. Load 10%-100%FS (this must be equal or larger than that for the second calibration point) weight o
6. Configuration Parameters Setup 1. When the scale is off, press and hold the ZERO/ON/OFF and PRINT/HOLD keys until ‘C0NF’ is shown, which indica
PS‐102 www.measuretek.net .7C8 2 Zero range for ZERO/ON/OFF key: 0 –power-on zero -point ±1%FS; 1 - power-on zero –point ±2%FS; 2 - power-on zer